Procedure
for performing Bubble point test.
o
Requirements:
§
Bubble point test assembly.
o
Procedures:
§
Remove the sterilized membrane filter assembly
in sterile area.
§
Attach sterile silicon tubing at both ends.
§
Fill 2 liters sterile water for injection in
bubble point test apparatus.
BUBBLE POINT TESTER
§
Attach upper end of silicon tube to the bubble
point test apparatus and lower end to another sterile glass flask.
§
Filter 2 liters of sterile distilled water
through membrane using Nitrogen pressure.
§
Remove the upper silicon tube and attach
Nitrogen tube to the upper end of membrane holder.
§
Gradually increase the Nitrogen pressure up to
25 PSI (1.5 lbs) and check for 30
seconds.
§
No bubble should come out from lower end of
lower tube (which is dipped in the sterile distilled water) and there should be
no fall in pressure,
Forms
and Records
o
Nil