PROCEDURE
FOR DETERMINATION OF THE FILLED VOLUME OF AMPOULES
AND VIALS.
Procedure
Precautions:
- Use protective hand gloves.
- Ensure that the limit of fill weight of the ampoule / vial is given in the respective BMR
Operation:
- Keep the weighing balance on the vibration free table.
- Switch on the weighing balance. Ensure that the pan of weighing balance is completely cleaned.
- Ensure weighing balance is showing ‘zero’ reading.
- Ensure that the weighing balance is calibrated.
- Keep the filled unsealed ampoule on the pan of weighing balance.
- Note down the position of the filled ampoule on the filling Machine.
- Remove all solution from the ampoule.
- Ensure that the solution is completely removed out.
- Again weigh the empty ampoule after removing solution.
- The weight of the solution would be the difference between the weight of the filled unsealed ampoule and empty ampoules.
- The weight of the solution should be within the range.
- Discard the ampoule after checking the volume.