Precautions And Instructions:
Ø
Ensure that the chamber of the autoclave is
cleaned properly.
Ø
Ensure that the main steam line pressure is
sufficient to operate the autoclave (i.e. minimum required steam pressure = 6.0
Kg/cm2 )
Ø
Always ensure that that if one door is open, the
other door is closed.
Ø
Always load the filled and sealed ampoules/vials
which are to be sterilize, from sterile end of the autoclave and always load
the general load from non sterile end of the autoclave.
Ø
Ensure that the circular chart is loaded in the
circular chart recorder.
Procedures:
Ø
Open the steam condensate valve and drain the
condensate from the steam line.
Ø
Open the door of the autoclave and load the
items to be sterilized (ampoules, vials or general load) from the appropriate
side (Sterile Side or Non Sterile Side) as per the loading pattern.
Ø
Close the door of the autoclave.
Ø
Switch on the circular chart recorder.
Ø
Open the steam inlet valve and allow the steam
to enter in the jacket.
Ø
Once the jacket becomes full (pressure = 1.5 Kg/cm2
), adjust the multi port valve to STER
position and allow the steam to enter from jacket to chamber. The drain temperature will increase up to
121° C.
Ø
Now, note down the “ster hold” start time.
Ø
Continue the “ster hold” as per the
sterilization parameters mentioned in the annexure - I
Ø
After completion of the “ster hold”, turn the
multi port valve to “Slow exhaust” position and simultaneously open the
compressed air valve (if applicable) slowly in such a way to maintain the
pressure of the chamber at 15 lbs psi.
Ø
Continue the cooling till the drain temperature
reaches to 70°C.
Ø
Now, stop the compressed air supply and ensure
the complete exhaust of the chamber.
Ø
Open the door of unloading side and take out the
entire sterilized load.
Forms and Records
Ø
Nil